• Skip to main content
  • Skip to secondary menu
  • Skip to primary sidebar
  • Skip to footer

Life in the Fast Lane • LITFL • Medical Blog

Emergency medicine and critical care medical education blog

  • MENU
  • ECG
  • CCC
  • Top 100
  • PODCASTS
  • JOBS
  • TOX
  • PART ONE
  • INTENSIVE
Home | CCC | Temporary Pacemaker Troubleshooting

Temporary Pacemaker Troubleshooting

by Dr Chris Nickson, last update March 25, 2019

Reviewed and revised 18 August 2014

OVERVIEW

Problems with pacing

  • output failure
  • failure to capture

Problems with sensing

  • oversensing
  • undersensing

Pacemaker syndromes

  • Cross-talk
  • Pacemaker syndrome
  • Pacemaker-mediated tachycardia
  • Sensor-induced tachycardia
  • Runaway pacemaker
  • Lead displacement dysrhythmia
  • Twiddler’s syndrome

GENERAL MANAGEMENT

Systematic approach is essential

  • review rhythm strip and 12 lead ECG
  • check integrity of circuit (start at patient -> pacing box): lead placement, polarity, integrity, tightly connected to correct port of pacing box (atrial/ventricular), battery, settings
  • check mode
  • check rate
  • check capture threshold (find threshold and double it for safety)
  • check sensitivity (normal = 2-5mV) – changes with position
  • fixes: change patient position, reverse bipolar pacing leads, convert to unipolar pacing, replace pacing equipment, return to OT for reinsertion of epicardial wires
  • back up plan in emergency: transcutaneous or tranvenous pacing, atropine, adrenaline, isoprenaline, ephedrine, electrolyte correction

FAILURE TO PACE DUE TO OUTPUT FAILURE

  • no electrical output at the pacing wire tips (pacing spikes absent on ECG)
  • causes: lead malfunction, unstable connection, insufficient power, cross-talk inhibition, oversensing (see below), apparent failure to pace
    ⇒ check power, battery and connections
    ⇒ increase output to maximum (20mA atrial and 25mA ventricular)
    ⇒ switch to an asynchronous mode to prevent oversensing (AOO, VOO)
    ⇒ connect the pacemaker directly to the pacing lead (occasionally the connecting wires may be faulty)
    ⇒ prepare for transcutaneous pacing
    ⇒ prepare for CPR and chronotropic drugs

FAILURE TO CAPTURE

  • visible pacing spikes are seen on ECG but no electrical capture on ECG or cardiac contraction seen in arterial line or SpO2 waveform
  • usually due to some specific mechanical problem (wires no longer connected to heart, wires not tightly connected to cable, cable not connected to correct port, output setting to low)
  • other causes: fibrosis at wire-myocardium interface, MI, electrolyte imbalance, post-defibrillation, drugs (flecanide, sotalol, betablockers, lignocaine, verapamil)
  • approach:
    ⇒ correct exacerbating causes
    ⇒ tight and confirm all external connections
    ⇒ increase output if possible
    ⇒ bipolar leads may be tried in reverse positions or can try convert to unipolar pacing
    ⇒ in bipolar leads, the negative electrodes develop fibrosis first -> use other electrode and plug into negative terminal and insert return electrode in the subcutaneous tissue (create unipolar circuit)
    ⇒ may need temporary transvenous wire

FAILURE TO SENSE

  • produces atrial pacing when not appropriate
  • due to specific setting of sensitivity (including AOO mode)
    ⇒ same mechanisms as failure to capture and pace
    ⇒ decrease absolute value of sensitivity (making it easier to inhibit)

OVERSENSING

  • Oversensing occurs when electrical signal are inappropriately recognised as native cardiac activity and pacing is inhibited
  • produces inappropriate/excessive inhibition of atrial pacing -> confuses pacemaker into thinking that there has been a return to spontaneous atrial activity
  • These inappropriate signals may be large P or T waves, skeletal muscle activity or lead contact problems
  • Abnormal signals may not be evident on ECG
  • Reduced pacemaker output / output failure may be seen on ECG monitoring if the patient contracts their rectus or pectoral muscles (due to oversensing of muscle activity)
  • usually due to settings on the pacemaker
    ⇒ increase absolute value of sensitivity (making it harder to inhibit)
  • in DDD external electrical impulses can also be misinterpreted as atrial activity causing pacemaker mediated tachycardia
    ⇒ increase sensitivity threshold or switch to an asynchronous mode (AOO, VOO)

CROSS TALK

  • in dual chamber pacing it is possible that the atrial pacemaker spike will be sensed by the ventricular wire and is misinterpreted as a ventricular depolarisation
    -> inhibits ventricular pacemaker output (ventricular standstill)
  • the opposite can happen as well
    ⇒ reduce sensitivity in atrial or ventricular channel
    ⇒ reduce mA delivered to the ventricular or pacing wire

PACEMAKER MEDIATED TACHYCARDIA

  • Also known as endless-loop tachycardia or pacemaker circus movement tachycardia
  • VDD or DDD pacing problem
  • can switch to VVI or DVI (but may lose AV synchrony)
  • Mechanisms:

(1) atrial sensing of a ventricular spike -> interpreted as an endogenous atrial depolarisation -> another ventricular impulse
⇒ use an atrial blanking period (now preset into box)

(2) retrograde conduction between ventricle and atrium through AV node or accessory pathway -> retrograde p waves being sensed as native atrial activity with subsequent ventricular pacing -> paced ventricular complex results in further retrograde conduction with retrograde p wave generation ->  ‘endless’ loop of periodicity -> re-entry tachycardia
⇒ adjustable post ventricular (pacing spike) atrial refractory period (PVARP) or slowing AV conduction e.g. adenosine or activation of magnet mode.

  • Results in a paced tachycardia with the maximum rate limited by the pacemaker programming
  • Newer pacemakers contain programmed algorithms designed to terminate PMT
  • May result in rate-related ischaemia in the presence of IHD

PACEMAKER SYNDROME

  • Caused by improper timing of atrial and ventricular contractions resulting in AV dyssynchrony and loss of atrial “kick”
  • Variety of clinical symptoms including fatigue, dizziness, palpations, pre-syncope
  • Associated decrease in systolic blood pressure > 20 mmHg during change from native rhythm to paced rhythm

TWIDDLER’S SYNDROME

  • Patient manipulation of the pulse generator (accidentally or deliberately)
  • The pacemaker rotates on its long axis, resulting in dislodgement of pacing leads
  • Can result in diaphragmatic or brachial plexus pacing (e.g. arm twitching) depending on extent of lead migration

LEAD DISPLACEMENT DYSRHYTHMIA

  • A dislodged pacing wire may float around inside the right ventricle intermittently “tickling” the myocardium and causing ventricular ectopics or runs of VT alternating with failure of capture.
  • If the paced QRS morphology changes from a LBBB pattern (indicating RV placement) to a RBBB pattern (indicating LV placement), this suggests that the electrode has eroded through the interventricular septum.
  • A chest x-ray will usually help to confirm the diagnosis.

RUNAWAY PACEMAKER

  • This potentially life-threatening malfunction of older-generation pacemakers is related to low battery voltage (e.g. overdue pacemaker replacement)
  • The pacemaker delivers paroxysms of pacing spikes at 2000 bpm, which may provoke ventricular fibrillation
  • Paradoxically, there may be failure to capture — causing bradycardia — because the pacing spikes are very low in amplitude (due to the depleted battery voltage) and because at very high rates the ventricle may become refractory to stimulation
  • Application of a magnet can be life saving but definitive treatment requires replacement of the pacemaker

SENSOR INDUCED TACHYCARDIA

  • Modern pacemakers are programmed to allow increased heart rates in response to physiological stimuli such as exercise, tachypnoea, hypercapnia or acidaemia
  • Sensors may “misfire” in the presence of distracting stimuli such as vibrations, loud noises, fever, limb movement, hyperventilation or electrocautery (e.g. during surgery)
  • This misfiring leads to pacing at an inappropriately fast rate
  • The ventricular rate cannot exceed the pacemaker’s upper rate limit (usually 160-180 bpm)
  • These will also usually terminate with application of a magnet

References and Links

LITFL

  • CCC — Pacemakers
  • ECG Library — Pacemaker malfunction
  • ECG Library — Pacemaker Rhythms – Normal Patterns

Medmastery

  • Types of pacemakers
  • Common pacemaker issues: Failure to capture
  • Common pacemaker issues: Undersensing
  • Common pacemaker issues: Oversensing
  • Troubleshooting with vectors
  • Course: Pacemaker Essentials

Journal articles

  • Reade MC. Temporary epicardial pacing after cardiac surgery: a practical review. Part 2: Selection of epicardial pacing modes and troubleshooting. Anaesthesia. 2007 Apr;62(4):364-73. Review. PubMed PMID: 17381573. [Free Full Text]

CCC 700 6

Critical Care

Compendium

…more CCC

Share this:

  • Facebook
  • Twitter
  • LinkedIn
  • Print

Related

About Dr Chris Nickson

An oslerphile emergency physician and intensivist suffering from a bad case of knowledge dipsosis. Key areas of interest include: the ED-ICU interface, toxicology, simulation and the free open-access meducation (FOAM) revolution. @Twitter | INTENSIVE| SMACC

Reader Interactions

Leave a Reply Cancel reply

This site uses Akismet to reduce spam. Learn how your comment data is processed.

Primary Sidebar

  • Twitter
  • Facebook
  • Vimeo
  • LinkedIn
  • Twitter
  • Twitter
  • RSS Feed
ECG Mastery

Recent Posts

Mastering Intensive Care 048 with Laura Rock

Road to Resus 1: Pre-hospital Resuscitation

Adult CXR Cases 002

Is climate crisis a medical emergency?

Pain scale

Footer

RSS FEED  LITFL posts by EMAIL or RSS

RSS FEED  LITFL Review by EMAIL or RSS

RSS FEED  FFFF by EMAIL or RSS

#FOAMed Medical Education Resources by LITFL is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License.

Based on a work at https://litfl.com

  • Twitter
  • Vimeo
  • Facebook
  • LinkedIn
  • Twitter
  • Twitter

Authors • Blog • Contact • Disclaimer

2018 Launch

  • 22,634,214 visitors

Copyright © 2019 · Powered by vocortex and iSimulate